AXIS

ATA error codes

13 error codes

Review the 13 error and warning messages from the AXIS ATA manufacturer manual, with meanings and fix steps. Device type: precision / analytical balance.

C-1 Self-test error (C-1) One of the startup self-tests (C-1) failed. If the message stays on for more than 1 minute, there is a fault. C-2 Self-test error (C-2) One of the startup self-tests (C-2) failed. If the message stays on for more than 1 minute, there is a fault. C-3 Self-test error (C-3) One of the startup self-tests (C-3) failed. If the message stays on for more than 1 minute, there is a fault. C-4 Self-test error (C-4) One of the startup self-tests (C-4) failed. If the message stays on for more than 1 minute, there is a fault. C-5 Self-test error (C-5) One of the startup self-tests (C-5) failed. If the message stays on for more than 1 minute, there is a fault. C-6 Self-test error (C-6) One of the startup self-tests (C-6) failed. If the message stays on for more than 1 minute, there is a fault. L Weighing pan missing The scale does not detect the weighing pan. The pan is not installed or the scale has mechanical damage. H Overload The load on the weighing pan exceeds the scale's weighing capacity, or the scale has mechanical damage. UnLOAd Remove the load from the weighing pan (unload) There is a load on the weighing pan at power-on, or a transport lock or foreign object is left under the pan. SErVICE Service required – mechanical damage Mechanical damage was detected in the scale's measuring system. Service is required. ------ Taring could not be completed The taring process could not be completed. Dashes remain on the display. Err t Operating temperature out of range The ambient temperature is outside the scale's permitted operating temperature range. Err C09 Internal calibration error The process was disturbed during internal calibration, or the calibration result is wrong.