A&D

MX50 error codes

15 error codes

Review the 15 error and warning messages from the A&D MX50 manufacturer manual, with meanings and fix steps. Device type: moisture analyzer.

CH no Self-check ınternal error CH no reports that the self-check function found an internal error. Cl PF Clock battery error Cl PF reports that the internal clock lost the date and time because of a battery problem. Cl Err Clock error Cl Err reports a fault in the internal clock circuit. CLOSE Heater cover open CLOSE reports that the heater cover is open when the self-check starts. error0 Internal error Error 0 reports an internal error in the moisture analyzer. error3 IC error Error 3 reports an integrated circuit (IC) error in the moisture analyzer electronics. error8 IC error Error 8 reports an integrated circuit (IC) error in the moisture analyzer electronics. error9 IC error Error 9 reports an integrated circuit (IC) error in the moisture analyzer electronics. Ht Err Temperature control error Ht Err reports that the heater temperature cannot be controlled correctly. t-UP Temperature calibration timeout t-UP reports that no key was pressed for five minutes during temperature calibration. E Overload E reports that the load on the sample pan exceeds the weighing capacity. -E Sample pan error -E reports that the reading is too low and the sample pan may not be in place. CAL E Calibration weight too heavy CAL E reports that the calibration weight was measured heavier than expected. -CAL E Calibration weight too light -CAL E reports that the calibration weight was measured lighter than expected. MEM FUL Result memory full MEM FUL reports that the stored measurement results have reached the upper limit.